ORP-TRA-600
ORP-TRA-600 online ORP transmitter is an accurate instrument used for industrial on-line PH measurement and PH controlling. The instrument is easy to install and use, operates safely, has a strong adaptability to the environment, and requires little maintenance. It widely used for the water treatment environmental protection, industrial waste water, chemical industry etc. The instrument has the following main characteristics: 1) Display measuring ORP value.. 2) High and Low limit relay control output, 3) Isolated 4-20mA current signal output or RS485 output.
Product Details
Contact Now
Product Details
Major Technical Specifications
Measurement range | -1000mV ~ +1000mV |
Resolution ratio | 0.2 mV |
Accuracy | ±0.2 mV |
Displaying mode | 3½ LCD |
Calibrating mode | One-point calibration (automatic identification of solutions) |
Manual temperature compensation | 0~120℃ |
Compression resistance of electrodes | 0~0.5MPa |
Measurement distance | standard configuration 5m(≤40m) |
Flow Velocity | ≤3m/s |
Pressure | ≤0.4Mpa |
Output | 4-20mA current or Modbus RS485 (optional) |
Control output | Maxi. and Mini. (constantly open) contact point |
Capacity of contact points | 3A/AC250V (with resistant loads) |
Power supply | DC-24V ,0.5A □; (Standard) DC-12V ,0.5A □; (Optional) |
Environmental conditions | Humidity: ≤85%/RH; Temp.: 0~50℃ |
External dimensions | 92*76*59mm (H x W x D) Installation mode: DIN Installation |
Wire Connection Details
Note: The standard version is a two-wire ORP electrode
A Terminals:
No. |
Terminal |
Terminal Content |
1 |
HI |
Constantly open high limit relay alarm output |
2 |
COM |
Relay common terminal |
3 |
LO |
Constantly open low limit relay alarm output |
4 |
NC |
Empty |
5 |
NC |
Empty |
6 |
INPUT |
ORP electrode signal input( Transparent wire) |
7 |
REF |
ORP electrode signal ground wire( Black) |
8 |
TEMP |
Empty |
9 |
TEMP |
Empty |
B Terminals
No. |
Terminal |
Terminal Content |
1 |
DC 24+ |
Power input + |
2 |
GND |
Equipment Ground wire |
3 |
DC 24- |
Power input - |
4 |
NC |
Empty |
5 |
485G |
RS 485 Signal Ground wire |
6 |
A |
RS485 Signal A |
7 |
B |
RS485 Signal B |
8 |
MA+ |
4-20mA current signal output + |
9 |
MA- |
4-20mA current signal output - |
Operation Of The Instrument
4.1 Soft key function
Soft key |
Key function |
Function description |
|
Menu Key |
Under Measuring status, Enter or exit related parameter “setting Menu” |
|
Right shift key |
Right Shift to select the digit of the parameter(Number flashes When selected); Up shift to browse Menu. |
|
Add shift key |
Change the selected parameter digital value and parameter shift |
|
Confirm key |
Confirm to select and enter into Menu and parameter save, |
4.2. Operation of the instrument
Under main interface,Press“ ”key to Menu interface , the detailed description as follows:
No. |
Menu |
Menu Content |
1 |
Un-T |
Measuring Unit: pH/mV shift to display |
2 |
T--C |
Temp Compensate Interface: (ORP can’t work) |
3 |
HC-- |
Relay high limit close value |
4 |
HO-- |
Relay high limit open value |
5 |
LC-- |
Relay low limit close value |
6 |
LO-- |
Relay low limit open value |
7 |
DE-- |
delay x second alarm |
8 |
4--A |
4mA corresponding PH value |
9 |
20-A |
20mA corresponding PH value |
10 |
ADD- |
RS-485 Address, default value 002 |
11 |
BAUD |
RS-485 Baud Speed, default value 9600 |
12 |
CAL- |
ORP Calibration Menu |
13 |
FAC 0 |
Restore factory settings for user |
14 |
FAC - |
Factory setting Menu, only for factory |
Note:After having set an option, you must press the button to save the data.
Calibration Of Instruments
The calibration takes the form of one-point calibration. The automatic calibration protection function work in the process of calibration, illegal calibration is not executed.
Before the calibration starts,make sure that you have compounded, the default calibration standard is 240mV, this value can be changed. Before the calibration starts, the electrodes must be submerged and activated in clean water or the measured solution for 5 minutes.
The calibration methods are as follows: when it is the in the measuring state,
Press the button and
until the instrument display “CAL” , and it enters the calibrating state.
Press the button , and the instrument will display “240mV” . Put the cleaned electrodes into the “240mV” buffer solution,
press the button , the controller display current buffer solution “240mV” value, when ORP value stable, and wait 3 to 5 minutes, press the button
to save.
l During calibration, if the controller display “ ERR” , which indicate calibration failure, need to re-calibrate, if more times calibration failure, pls change the ORP electrode, or contact with the supplier.
Installation Of Measuring Electrodes
The measuring electrodes are compound electrodes with plastic shells and 3/4 " NPT threaded joints, Please install strictly according to the following diagrams:
3/4"BSP threaded joints.
Points for attention:
1) The electrodes shall be installed where the flow speed is steady and air bubs seldom gather or backwater which may lead to the inaccuracy.
2) The electrode shall be submerged deep into flowing water.
3)The cable collecting PH signals which is a form of weak potential signals shall be separately wired and shall not share a protective pipe with power supply wires and control wires, so as to avoid interference.
Measuring electrodes maintenance:
1) Avoid putting the electrodes in the dry air. When the electrodes are not in use, clean them with water and then put them back into the liquid- containing case with 1 mol KCL,or put the electrodes into a container with 1 mol KCL in it.
2) The electrodes shall be cleaned regularly. If the glass bulbs of the electrodes and Polytetrafluoroethylene (PTFE) are contacted and polluted, clean them with the following agents:
① use surface active agent to get rid of grease or greasy things.
② use 10% hydrochloric acid solution to get rid of calcium precipitates or metal hydroxides.
③ use 10% hydrochloric acid solution to get rid of sulphide precipitates. Use the mixture of 10% hydrochloric acid solution and pepsins to get rid of protein adhesives.
Maintenance Of Instrument
1) The measuring electrodes are accurate parts and shall not be taken out of the measuring room unless it is necessary to do so.The ORP electrode shall be cleaned Regularly.
2) Measuring cables are special cables which shall not be replaced by other types of cables. Unqualified cables may cause big errors.
3) If the measuring electrodes are damaged or losing efficacy, Purchas and replace them with the same type of ORP electrodes, and calibrate them according to the fifth chapter of this manual.
4) The instrument is composed of accurate integrated circuits and electronic components and therefore shall be protected from humidity.
The Instrument As A Whole Set
ORP transmitter |
1pc |
ORP Sensor |
1pc |
Fixing clamp |
1 pair |
Operation manual |
1 copy |
Get in Touch
If you are interested in our products, you can choose to leave your information here, and we will be in touch with you shortly.
FAQS
How to use ORP - TRA - 600 Operation Manual?
How to carry out the calibration operation in the manual?
How does the manual guide parameter setting work?
What suggestions does the manual have for troubleshooting?
Related Products
Related News
Apr.21,2025
Understanding Turbidity Meter Types: Which One Is Right for Your Application?
Monitoring turbidity—an indicator of water clarity—is vital for applications ranging from drinking water treatment to environmental monitoring.
Apr.21,2025
Understanding Total Dissolved Solids
Water may look clear, but that doesn’t mean it's pure. Hidden within every glass can be a range of minerals, salts, metals, and organic substances collectively known as total dissolved solids.
Apr.21,2025
Transmitter 4 20mA Technology
In the world of industrial automation and process control, accurate signal transmission is everything.
Apr.21,2025
Smart RO System Controller
In today’s world, clean water isn’t just a luxury—it’s a necessity. Whether in residential settings or industrial applications, ensuring a consistent supply of high-quality water has become vital.
Apr.21,2025
Residual Chlorine Meter is Essential
Chlorination is one of the most widely used methods for disinfecting water, whether it’s for municipal supply, swimming pools, or industrial processes.
Apr.21,2025
Essential Water Sampler Types
When it comes to environmental analysis, wastewater treatment, or drinking water inspection, water sampler types play a crucial role in collecting representative water samples for laboratory analysis.
Apr.21,2025
Dissolved Oxygen Meter
Water may look clean, but what’s happening on the molecular level often tells a different story.
Apr.21,2025
Different Types of Sensors in Industrial Automation
In the rapidly evolving age of smart manufacturing and Industry 4.0, the different types of sensors deployed across production facilities are nothing short of transformative.
If you are interested in our products, you can choose to leave your information here, and we will be in touch with you shortly.