WS 9100
MODEL 9100 Series Portable Automatic Water Sampler has been designed to offer the user a wide range of options in a specialized environmental monitoring instrument. The unit uses a 12 VDC power source and can be operated using a 230/ 120 VAC to 12 VDC power supply, 12 VDC battery or 12 VDC battery\solar panel package for remote location sampling. The 2100 Series incorporates a temperature- controlled sample storage chamber that utilizes both refrigeration and heating to maintain sample temperature requirements. There is no need for ice or to stop monitoring when outdoor temperatures go below freezing. It uses a peristaltic pump design for liquid draw. The unit is controlled by a single- microchip computer. Fully programmable, the user can set sampling time interval, sampling quantity and sampling mode, as well as other application parameters according to their requirements. The 2100 series can be equipped with a flow sensor to accomplish volume-based sampling as well and time-based sampling.
Product Details
Contact Now
Product Details
Specifications
1.Sampling Method:
* Composite sampling:
Sampling delay time: Any setting of 1-9999 minutes with increments of 1minute.
Sampling interval time: Any setting of 1-9999 minutes with increments of 1minute. Sampling volumes:50~7500 ml with increment in 1 ml.
Number of mixed samples:1-200
*Flowmeter control sampling:Dedicated flowmeter control sampling *Sampling with Flow Sensor with Pulse output. Control 1~ 9999 pulses.
2. Characteristic of Peristaltic Pump:3.7L/min.
3. Sampling Error:±5%
4. Repeat Sampling Accuracy:±5ml
5. Vertical Head:8m
6. Horizontal Suction Head: 50m. Sampling tube has automatic air purge function.
7. Sampling Bottle:1 x 7500ml Polyethylene bottle
8. System clock time control error: △1≤0.05% and △12≤3S
9. Air tightness of pipeline system: ≤-0.05Mpa
10. Mean Time Between Failure (MTBF): ≥1440h/time
11. Insulation Resistance: >20MΩ
12. Software lock, it can protect programs from modification.
13. It can preset and save maximum 10 programs according to needs.
14. Sampler can auto-record and save samples. The stored data will not be lost when power off, so you can check it at any time. Users can perform data analysis based on these data and water sample test results.
15. Working temperature:-25℃ to +50℃
16. Power: 12V DC, 80W
17. Size:580mm(L) x 320mm(W) x 520mm(H); Weight: 15Kg
Installation Precautions
1 . When placing the sampling bottle, be sure to press the sampling bottle to the bottom to prevent dispenser from colliding with the top of sampling bottle and not working properly.
2 . The pump tube should be checked regularly and replaced if damaged. Leave a gap of 2mm between the
pump tube and the pump shell. After pump tube is replaced, press ↑ (up) to make the pump rotate forward, or press (down) to make the pump anti-clockwise running. Let the pump rotate for a while, until the pump runs normally without jamming, and the pump tube is installed successfully.
3 . If the sampler is abnormal, please cut off the power immediately and notify the manufacturer to repair it. 4 . If the pumping accuracy does not meet the setting parameter, user can adjust the length of pumping pipe to meet the requirements. For more details, please refer to pumping parameter settings.
Programming Guide
After system is on, interface shows:
MODEL 9100 Automatic Water Samplers MODEL 9100 2020-02-20 20:05:46 |
With the above display Press ↑(up) to make the pump rotate forward. Press ↓ (down) to make the pump rotate in the reverse direction.
1. Manual Sampling
When in startup display, press→(right) to enter Manual sampling mode:
MODEL 9100 Automatic Water Samplers
Manual sampling Set 00100ml (50-7500ml)
Press Enter to confirm Press M to cancel |
Press ←(left) or → (right) to move cursor. Press ↑(up) or ↓(down) to increase or decrease number by 1. After setting the desired volume, press Enter ↵ (Enter) to save the setting. Press (M) to cancel. Use the same procedure for each parameter.
2. Automatic Sampling
Press ↵ (Enter under start-up interface to enter automatic sampling mode:
MODEL 9100 Automatic Water Samplers Auto-sampling Please enter program number 01 Press Enter to confirm Press M to cancel |
Select required program number then press ↵ (Enter) go to Auto-sample model. Press M button to cancel.
3. Automatic Sampling Parameter Setting
Menu: Press M to enter main menu (Press M again go back to start-up interface).
Menu |
Sample Setting |
Flowmeter System Setting Factory Setting About Exit |
Select “Sample Setting”
Sample Settings |
Set sample |
Query Sampling history Exit |
Press ↵(Enter) to sample parameter setting interface:
Set Sample Parameter Sun 1 Mon 1 Tues 1 Wed 1 Thur 1 Fri 1 Sat 1 Start time 08:00 End time 20:00 Interval time 0000 Min ( 1-9999) Volume 00200 ml (50-7500) Mixed number 050 ( 1-200) Sample number 01
Press Enter to confirm Press M to cancel |
To cancel setting, press M, going back to “ Menu” . To save the setting press ↵(Enter). As below:
Setting Confirm
Number 01 Save or cancel?
Press Enter to save Pres M to cancel |
The Sampler can save and install up to 10 preset programs. After setting, select “Query” to find the preset programs parameters.
Sample Setting Sampling setting |
Query |
Sampling history Exit |
Query |
Sun 1 Mon 1 Tues 1 Wed 1 Thur 1 Fri 1 Sat 1 Start time 08:00 End time 20:00 Interval time 0030 Min ( 1-9999) Volume 00200 ml (50-7500) Mixed number 050 ( 1-200) Sample number 01
Press Enter to confirm Press M to cancel |
4. Folowmeter Setting
Press M on start-up interface, then you can go into the Folowmeter interface, as following:
Flowmeter No flowmeter |
4-20mA |
Pulse flowmeter Flow data Exit |
Now, please choose according to your needs. such as select the 4-20mA and confirm.
4-20mA
Select 4-20 mA mode Press Enter to confirm Press M to cancel |
4mA 20mA Start value |
4-20mA = 0000000 = 0000100 = 0000100 |
Press Enter to confirm Press M to cancel |
Set here how much 4mA represents and how much 20mA represents. And what is the starting value. such as select the Pulse flowmeter and confirm.
Set number of pulses Current: 0000001
Start value: 0000001
Press Enter to confirm Press M to cancel |
How many pulses can you set to start.
Short circuit between pulse 6 and ground 7 is a pulse.
Go back to the main menu, Press Enter from Star-up interface, going into the automatic sampling interface, as the following:
MODEL 9100 Automatic Water Samplers Auto-sampling Please enter program number 01 Press Enter to confirm Press M to cancel |
Now our sampler goes into automatic sampling status.
5. Time Setting
At the Start-up display → System setting as below:
System settings |
Time Setting |
Communication Setting Peristaltic Pump Parameter setting Password Setting Blacklight Exit |
Press ↵(Enter ) to Time Setting:
Time Setting
2023-01-01 20:35:58
Press Enter to confirm Press M to cancel |
6. Suction Tube Length Setting
Go into Suction Tube Length Setting interface, select the normal length.
Suction Tube Length Setting Current: 0236 in (40-2000) Set: 0236 in Press Enter to confirm Press M to cancel |
It the actual tube is longer or shorter than the pump tube provided, you should update the data.
7. Manual Calibration
Set Correction Factors Current: 0650 Set: 0000 Press Enter to confirm Press M to cancel |
How to set the Calibration:
1st step: Set up the length of suction tube.
2nd step: Press ↑(up) on the start-up interface to make the pump rotate forward. If there’s water is being dispensed the set up was successful; Otherwise, continue to press ↑(up) until water is dispensed.
3rd step: Repeat the step 2 step 3 or 4 times. When the water output amount is stable, you can measure the amount and input the number to “ Setting Coefficient” . Press ↵ (Enter) to save.
8. Number of Air Purges
Setting Number of Purges Before sampling: 1 Aftre sampling: 1 Set before sampling: 1 1Set after sampling: 1
Press Enter to confirm Press M to cancel |
The number of air purges indicates the number of backflushes before each sampling
9. Password Setting
Password Settings |
Password set/cancel |
Change password Exit |
As the above shows √ means password has been enabled. Press ↵ (Enter) to stop password enable state.
Original
Password:123456. If you want to change password, goes into the “Change password” interface:
Please enter new Password 0 0 0 0 0 |
10.Blacklight Time
Blacklight Time Setting Current: 300 (5-300Sec). Set : 300 Press Enter to confirm Press M to cancel |
11. Refrigerator
A. Connect to the power and press ON/OFF key to run on. Now the “Power” light is on. The digital display will indicate the current temperature inside the chamber.
B. If the power is off or disconnected, the refrigerator will shut down automatically. But it will operate in the mode set before when power returns.
C. Press △UP or ▽DOWN can set the temperature.
D. When continuously press “SET” until the display shows “ECO” or “HH”, you can set the refrigerator to run in ECO energy-saving mode or MAX(HH) speed mode. The corresponding indicator light will be on.
E. Under “ECO” mode, the compressor is running at low speed to save energy. When working in “MAX”
mode, the compressor is running at high speed, and the power consumption will increase. It is recommended to start the MAX mode when rapid colling is required or the freezing temperature needs to be as low as -
20°C.
12. About
Press M to start-up interface, “About” as:
About Automatic Water Sampler 6rd Version 2023/04/18 Hardware: Product: Use: |
13 Temperature Controller:
Heating Mode Setting
Key Point : Start up temperature < Stop temperature ( The program automatically turn on the heating mode)
Set up:
1. Long press the▲up botton until the screen flashes. Then press the ▲ up or ▼down botton to set up the Start up temperature.
2. long press the ▼ down botton until the screen flashes. Then press the ▲ up or ▼ down botton to set up the Stop temperature.
For example:
How should I set it up if I want to stop heating at 4 ℃ , and start heating again when temperature drops to 0℃
1st :Make sure it is in heating mode, Start up temperature < Stop temperature
2nd :long press the ▲ up botton until the screen flashes. Set Start tempreture to 0℃.
3rd: long press the down▼ botton until the screen flashes. Set Stop tempreture to 6℃
Terminal diagram
RS-232 |
Ground |
Flowmeter |
Pulse |
Ground |
|||
1 |
2 |
3 |
4 |
5 |
6 |
7 |
8 |
Input |
Output |
4-20 ma - |
4-20 ma + |
+16 VDC |
+ |
- |
- |
Packing List
Enclosure |
1 |
Pumping Pipe |
1 |
Pumping pipe with filter |
1x6m |
Polycarbonate Sampling Bottle |
1x7500ml |
Operation Manual |
1 |
Certificate |
1 |
DC power cord |
1 |
AC power adapter |
1 |
Get in Touch
If you are interested in our products, you can choose to leave your information here, and we will be in touch with you shortly.
FAQS
What is the purpose of the 9100 series sampler?
How do I install a 9100 Series Sampler?
How do I operate the 9100 Series Sampler?
What's wrong with the 9100 series sampler?
Related Products
Related News
Apr.21,2025
Understanding Turbidity Meter Types: Which One Is Right for Your Application?
Monitoring turbidity—an indicator of water clarity—is vital for applications ranging from drinking water treatment to environmental monitoring.
Apr.21,2025
Understanding Total Dissolved Solids
Water may look clear, but that doesn’t mean it's pure. Hidden within every glass can be a range of minerals, salts, metals, and organic substances collectively known as total dissolved solids.
Apr.21,2025
Transmitter 4 20mA Technology
In the world of industrial automation and process control, accurate signal transmission is everything.
Apr.21,2025
Smart RO System Controller
In today’s world, clean water isn’t just a luxury—it’s a necessity. Whether in residential settings or industrial applications, ensuring a consistent supply of high-quality water has become vital.
Apr.21,2025
Residual Chlorine Meter is Essential
Chlorination is one of the most widely used methods for disinfecting water, whether it’s for municipal supply, swimming pools, or industrial processes.
Apr.21,2025
Essential Water Sampler Types
When it comes to environmental analysis, wastewater treatment, or drinking water inspection, water sampler types play a crucial role in collecting representative water samples for laboratory analysis.
Apr.21,2025
Dissolved Oxygen Meter
Water may look clean, but what’s happening on the molecular level often tells a different story.
Apr.21,2025
Different Types of Sensors in Industrial Automation
In the rapidly evolving age of smart manufacturing and Industry 4.0, the different types of sensors deployed across production facilities are nothing short of transformative.
If you are interested in our products, you can choose to leave your information here, and we will be in touch with you shortly.